Barco MFCD 1219 (TS) Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Telewizory i monitory Barco MFCD 1219 (TS). Barco MFCD 1219 (TS) User manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - MFCD 1219 (T S)

User ManualMFCD 1219 (T S)

Strona 2

Preface10This apparatus conforms to:CE (LVD 73/23/EEC), IEC 60950-1, UL 60950-1, CAN/CSA C22.2 No. 60950-1 (c-UL), EN 60950-1 DEMKO, CCC GB4943-1995Na

Strona 3 - Copyright notice

Preface11saver is not activated), change the image content regularly to avoid image retention of the static elements.3. Understand pixel technologyLCD

Strona 4 - Table of Contents

Overview12OverviewIntroductionThe MFCD 1219 (TS), BARCO’s full color 19” LCD display, guarantees perfect image quality in medical imaging applications

Strona 5

Overview13Touch panelThe MFCD 1219 TS is equipped with a touch panel. Instructions on how to install and use the touch panel driver and controls can b

Strona 6

Overview14Controls and connectors1. 12V DC power input2. DVI input (DVI-D supported only)3. Analog video input4. USB port for touch panel connection t

Strona 7 - Safety Instructions

Installation15InstallationPrecautions• Keep your original packaging. It is designed for this display and is the ideal protection during transport.• Av

Strona 8

Installation162. Firmly press down the push-button (9) with one hand while carefully lifting the panel up with the other hand.3. Remove the panel from

Strona 9

Installation17Please make sure that the four shafts are securely placed in the notches of the mounting plate.2. To secure the panel, slide forward the

Strona 10

Installation18To connect the video & sync signals:1. Connect one end of the appropriate video cable to the corresponding video input of the monito

Strona 11

Installation19Note: The recommended resolution for best image quality is 1280x1024 at 60 Hz.The microcontroller in the display checks if the connected

Strona 12 - Overview

2(This page intentionally left blank.)

Strona 13 - Package contents

Installation20Install the softwareMediCal QAWeb Agent software:Insert the CD-ROM in the CD-ROM drive. The CD-ROM starts automatically. Please follow t

Strona 14 - Controls and connectors

On-screen display (OSD)21On-screen display (OSD)How to use the OSDThe OSD is a window on your screen that allows you to view and change display settin

Strona 15 - Installation

On-screen display (OSD)22b) Press the Menu key (3) to enter the new value and return to the menu.5. The current settings are saved automatically when

Strona 16

On-screen display (OSD)23Brightness / Contrast menuFunction DescriptionBrightness To adjust the overall image and background brightness level. This fu

Strona 17 - Connecting the signals

On-screen display (OSD)24Color MenuFunction DescriptionNative Set the image color temperature to the LCD panel’s native, uncorrected color temperature

Strona 18 - Starting up

On-screen display (OSD)25Geometry menuThis menu is available only when an analog video signal is selected.Function DescriptionAuto adjust-mentSelects

Strona 19

On-screen display (OSD)26Settings menuNote: The DPMS system will power down the display when the connected computer is left idle for a certain time. B

Strona 20 - Install the software

On-screen display (OSD)27OSD Settings submenuTo unlock the Controls, proceed as follows:1. With one hand, press and hold the < (left) and the >

Strona 21 - On-screen display (OSD)

On-screen display (OSD)28Information menuFunction DescriptionSerial No. Indicated the display serial number.FW version Displays the current internal f

Strona 22 - Description of the menu items

Maintenance29MaintenancePanelTake care not to damage or scratch the panel. Clean with a soft woolen or cotton towel. Use a watery solution or a mild c

Strona 23 - Brightness / Contrast menu

3Copyright noticeThis document is copyrighted. All rights are reserved. Nor this document, nor any part of it, may be reproduced or copied in any fo

Strona 24 - Color Menu

Appendix A: Possible resolutions30Appendix A: Possible resolutionsThe following modes are preset as factory defaults.ResolutionPixelfreq.(MHz)Hor.freq

Strona 25 - Geometry menu

Appendix B: Technical specifications31Appendix B: Technical specificationsPicture panel 19-inch diagonal viewable screen TFT (thin film transistor) ac

Strona 26 - Settings menu

Appendix B: Technical specifications32NOTICE:DUE TO OUR POLICY OF CONTINUOUS PRODUCT IMPROVEMENT, THE ABOVE SPECIFICATIONS ARE SUBJECTED TO CHANGE WIT

Strona 27

Appendix B: Technical specifications33(This page intentionally left blank.)

Strona 28 - Information menu

Appendix B: Technical specifications34(This page intentionally left blank.)

Strona 29 - Maintenance

Appendix B: Technical specifications35(This page intentionally left blank.)

Strona 30

Barco nv Medical ImagingPresident Kennedypark 35B-8500 Kortrijk, Belgiumwww.barco.comK5904239-00August 2010

Strona 31

Table of Contents4Table of ContentsPreface ... 5Environm

Strona 32

Preface5PrefaceNoticeAlthough every attempt has been made to achieve technical accuracy in this document, we assume no responsibility for errors that

Strona 33

Preface6reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or mo

Strona 34

Preface7Environmental informationDisposal InformationThe lamps inside the display contain mercury. Do not throw the display in the trash. Dispose of i

Strona 35

Preface8Electrical shockType of protection (electrical):Class I equipmentDegree of safety (flammable anesthetic mixture):Equipment not suitable for us

Strona 36 - August 2010

Preface9Water and moisture Never expose the display to rain or moisture. Never use the display near water - e.g. near a bathtub, washbasin, swimming p

Komentarze do niniejszej Instrukcji

Brak uwag