System manual MeDis 5MP1HM 10(This page intentionally left blank)Introduction
System manual MeDis 5MP1HM 100To control contrast manually1 Turn the control knob to select Contrast in the OSD main menu.2 Press the control knob sho
System manual MeDis 5MP1HM 101 STATUS INFORMATIONDisplay Name: MGD 521MKIISerial Number: 5114147Mode Name: New modeMode H-frequenc
System manual MeDis 5MP1HM 102Software Version:The internal software version.Run time:The total operating time since production, including burn-in tim
System manual MeDis 5MP1HM 103ADVANCED SETTINGS ANDADJUSTMENTS(FOR SERVICE STAFF ONLY)Important:The functions and controls described in this chapter c
System manual MeDis 5MP1HM 104Explanation of the Settings menu itemsLUC:LUC is short for Luminance Uniformity Control. When switched on,the internal w
System manual MeDis 5MP1HM 105 ADJUSTMENTS- Geometry- Focus- ALC- Luminance uniformity- Factory reset- ExitAdjustments menuAdjustmentsa) How to make
System manual MeDis 5MP1HM 106WidthHorizontal positionHeightVertical positionBlanking - Blanking leftBlanking - Blanking rightBlanking - Blanking topD
System manual MeDis 5MP1HM 107Interlacing - Pre correctionIn most cases you do not need to change this adjustment.You may need to increase this adjust
System manual MeDis 5MP1HM 108SH Compensation: SH PositionWith SH Compensation on, we advise to use the SH Position controlinstead of the Horizontal P
System manual MeDis 5MP1HM 109Advanced geometry - Corner top rightAdvanced geometry - Corner bottom leftAdvanced geometry - Corner bottom rightAdvance
System manual MeDis 5MP1HM 11DISPLAYCONTROLLERINSTALLATION
System manual MeDis 5MP1HM 110Advanced geometry - Advanced bowing & skewing -Horizontal bowing leftAdvanced geometry - Advanced bowing & skewi
System manual MeDis 5MP1HM 111FocusStatic focusAdjust Static focus to obtain an image as sharp as possible in thecenter.Focus zoneYou can adjust the f
System manual MeDis 5MP1HM 112Luminance uniformityThis menu allows you to calibrate the display, including theLuminance Uniformity circuit.Proceed as
System manual MeDis 5MP1HM 113MAINTENANCEPicture tubeThe glass panel of the picture tube is handled with a special coating.Take care not to damage or
System manual MeDis 5MP1HM 114TROUBLESHOOTINGThere appears no image on the picture tube, the green LED atthe front is out Check if the power cord is
System manual MeDis 5MP1HM 115APPENDIX A:BACKGROUND INFORMATIONLuminance Uniformity CorrectionA characteristic (or limitation) of every picture tube (
System manual MeDis 5MP1HM 116At last, the imaging board delivers no sync signals at all, and themicro-controller switches off all but one power suppl
System manual MeDis 5MP1HM 117Picture tube:Faceplate transmission: 32%Faceplate type: ARC panelImage representation: portraitPhosphor: P45 (standard)
System manual MeDis 5MP1HM 12(This page intentionally left blank.)Display Controller installation
System manual MeDis 5MP1HM 13BarcoMed 5MP1HM Product OverviewThe BarcoMed 5MP1HM Display Controller delivers a quality 2048 x2560 pixel image with 102
System manual MeDis 5MP1HM 14(This page intentionally left blank.)Display Controller installation
System manual MeDis 5MP1HM 15HARDWARE INSTALLATIONNotice: Wear a protective ESD strap during installation or handling ofthe controller. Electrostatic
System manual MeDis 5MP1HM 16should now remove any third party VGA controller from your system ordisable the integrated VGA controller.If you decide t
System manual MeDis 5MP1HM 17Installing the BarcoMed 5MP1HM Display ControllerInstall the 5MP1HM into your computer following these steps:1. Turn off
System manual MeDis 5MP1HM 18Figure 4: The Video Outputs.:1DIVyalpsidtsrifehtrofrotcennoc)neerg(oediVehttcennoC.ereh:SHts1ehtrofrotcennoc)kcalB(cnySla
System manual MeDis 5MP1HM 19DISPLAYINSTALLATION
B410046 / 00July 2003© 2003 BARCO NV. All rights reservedBarcoViewTheodoor Sevenslaan 1068500 Kortrijk, BelgiumPhone: +32(0)56 23 32 44Fax: +32(0)56 2
System manual MeDis 5MP1HM 20(This page intentionally left blank.)Display installation
System manual MeDis 5MP1HM 21DISPLAY INSTALLATIONExampleAn example of a correct set-up is shown in figure 5.Please take a good look at the figures, an
System manual MeDis 5MP1HM 22Placing and configuring the displays1 The last part of this system manual is a copy of the display usermanual. Follow the
System manual MeDis 5MP1HM 23Connection of the Remote cablesNote: One of the remote cables is packed inside the MediCal box.1 The Remote cables that c
System manual MeDis 5MP1HM 24(This page intentionally left blank.)Display installation
System manual MeDis 5MP1HM 25BARCOMED DRIVERAND SOFTWAREINSTALLATION
System manual MeDis 5MP1HM 26(This page intentionally left blank.)BarcoMed driver and software installation
System manual MeDis 5MP1HM 27Planning Your Barcomed Windows 2000InstallationBecause of Windows 2000s Plug and Play features, it is veryimportant to c
System manual MeDis 5MP1HM 28If you choose to install multiple types of BarcoMed Imaging Boards inyour computer it is important to carefully note whic
System manual MeDis 5MP1HM 29Installing the BarcoMed 5MP2 AURA, 5MP2,5MP1HM, 5MP1H or 4MP2 Windows 2000SoftwareSPECIAL NOTE: Prior to installing the B
System manual MeDis 5MP1HM 3CONTENTSIntroduction ...
System manual MeDis 5MP1HM 30contents of your BarcoMed Software CD and double click on thefile: Setup.exe to start the wizard.The BarcoMed Product I
System manual MeDis 5MP1HM 31Figure 7Driver Installation7. Click Next on the Display Driver Wizards Welcome Screen tocontinue or Cancel to exit t
System manual MeDis 5MP1HM 32Figure 8Click OK to continue. If you dont want to replace the exitingdriver, click OK and then click Cancel.10. If
System manual MeDis 5MP1HM 33Figure 10: 5MP2 AURADualView EnabledIf you wish to enable DualView check the checkbox next toEnable the DualView devi
System manual MeDis 5MP1HM 34Figure 1114. When the Wizard tells you that it has successfully installed theselected driver, click Finish. The wizard
System manual MeDis 5MP1HM 35to enable DualView, you will need to reinstall the drivers with theEnable the DualView device? checkbox checked.Figure
System manual MeDis 5MP1HM 36Figure 1317. The Setup Type page will now appear (figure 14). ChooseTypical to install the default schemes with Engli
System manual MeDis 5MP1HM 37During installation the wizard will display a progress screen.19. When the InstallShield Wizard Completed page appears,
System manual MeDis 5MP1HM 38Setting the Resolution of Your 5MP1HM1. To set the resolution for your high resolution display(s), right clickon the des
System manual MeDis 5MP1HM 39snoituloseRelpmaSrellortnoCnoitarugifnoCyalpsiDnoituloseRpotkseDlautriVnoituloseRARUA2PM5enOanisyalpsid2worelgnis0652x840
Driver Installation ... 31BARCO DPMS Screen Saver Installation ...
System manual MeDis 5MP1HM 40Figure 16: The Windows 2000 Display Properties Control Panel Settings Tabafter the drivers have been installed and the re
System manual MeDis 5MP1HM 41Planning Your Barcomed Windows XPInstallationBecause of Windows XPs Plug and Play features, it is very importantto caref
System manual MeDis 5MP1HM 42If you have previously installed drivers for the BarcoMed DisplayControllers in your Windows XP System, Windows XP should
System manual MeDis 5MP1HM 43Installing the BarcoMed 5MP2 AURA, 5MP2,5MP1HM, 5MP1H or 4MP2 Windows XP SoftwareSPECIAL NOTE: Prior to installing the Ba
System manual MeDis 5MP1HM 447. Insert your BarcoMed Software CD into your computers CD drive.The BarcoMed Product Installation Wizard should start
System manual MeDis 5MP1HM 45Note: You can install the driver for only one type of BarcoMeddevice at a time. If you have multiple types of BarcoMed de
System manual MeDis 5MP1HM 46Figure 19: 5MP2 AURADualView EnabledFigure 20: 5MP2 AURADualView DisabledBarcoMed driver and software installation
System manual MeDis 5MP1HM 47If DualView is not enabled, you will see a single virtual screen inthe display properties dialog for both heads of a Barc
System manual MeDis 5MP1HM 48Note: Clicking Cancel will also return you to the BarcoMedProduct Install Wizard, but will not delete the Barco Display
System manual MeDis 5MP1HM 49Figure 2320. The Setup Type page will now appear (figure 24). ChooseTypical to install the default schemes with Engli
Work Schedule ... 70Begin Work Day ...
System manual MeDis 5MP1HM 5021. The Ready to Install Program page will now appear, clickInstall to install DPMS or Back if you wish to change a
System manual MeDis 5MP1HM 51Setting the Resolution of Your 5MP1HMOnce you have completed the installation of the display driver, youneed to set the r
System manual MeDis 5MP1HM 52Repeat steps 2 - 7 for each aditional BarcoMed 5MP1HM displaycontrolled installed in your system.If you are using a Quad-
System manual MeDis 5MP1HM 53Reinstalling or Updating your BarcoMed 5MP2 AURA, 5MP2,5MP1HM, 5MP1H or 4MP2 DriversTo reinstall or update only the Barco
System manual MeDis 5MP1HM 54(This page intentionally left blank.)BarcoMed driver and software installation
System manual MeDis 5MP1HM 55BARCOMEDCONTROLLERTOOLS
System manual MeDis 5MP1HM 56(This page intentionally left blank.)BarcoMed controller tools
System manual MeDis 5MP1HM 57BARCOMED DRIVER TABIntroductionAfter the BarcoMed Windows Display Controller Driver is installed, anew Display Properties
System manual MeDis 5MP1HM 58Figure 28: BarcoMed Driver Tab under Windows 2000Figure 29: BarcoMed Driver Tab under Windows XPBarcoMed controller tools
System manual MeDis 5MP1HM 59StatusDisplays the current BarcoMed Display Controller, driver, and thecurrently selected display resolution.Graphics Boa
Important considerations ... 97Operation: User controls ...
System manual MeDis 5MP1HM 60index 10 to 245 contain the missing gray values so that the palettehas the full 256 gray values within it.Please note tha
System manual MeDis 5MP1HM 61application requires this Static Gray Palette with NO System colorsmode, contact your application provider.Drawing Mode
System manual MeDis 5MP1HM 62one row option from the pull-down list. This would cause theWindows virtual desktop to be resized to fit on the three mo
System manual MeDis 5MP1HM 63BARCOMED HARDWARE TABIntroductionThe BarcoMed Hardware Tab is used for gathering information aboutBarcoMed Display Contro
System manual MeDis 5MP1HM 64To access the BarcoMed Hardware Tab do the following:1. Open the Display Properties Control Panel by right clicking on
System manual MeDis 5MP1HM 65Figure 31: BarcoMed Hardware Tab underWindows 2000 and Windows XPUsing The BarcoMed Hardware TabDeviceDisplays the curren
System manual MeDis 5MP1HM 66PCI InformationDevice ID: Displays the devices PCI Device ID number.Vendor ID: Displays the device manufacturers PCI Ve
System manual MeDis 5MP1HM 67update the firmware stored in the ROM of the currently selectedBarcoMed Display Controller. The BarcoMed Hardware Configu
System manual MeDis 5MP1HM 68(This page intentionally left blank.)BarcoMed controller tools
System manual MeDis 5MP1HM 69BARCO DPMS SCREEN SAVEROVERVIEWIntroductionThe BARCO DPMS (Display Power Management Signaling) ScreenSaver allows the use
System manual MeDis 5MP1HM 7INTRODUCTION
System manual MeDis 5MP1HM 70BARCO DPMS Screen Saver OptionsFrom the BARCO DPMS Screen Saver Control Panel you canchange many DPMS screen saver elem
System manual MeDis 5MP1HM 71Screen Saver During Work DayOn CRT based displays, such as the Barco MGD series, a regularscreen saver can be invoked d
System manual MeDis 5MP1HM 72etatSSMPD sgnivaSrewoProtinoMemiTyrevoceRnOenoNA/Nyb-dnatSlaminiMtrohSdnepsuSlaitnatsbuSregnoLffOmumixaMtsegnoLMonitor Se
System manual MeDis 5MP1HM 73privileges. If a user who does not have administrator privileges tries tosave a scheme using an existing default scheme n
System manual MeDis 5MP1HM 74Figure 32: Windows NT 4.0 Screen Saver TabBarcoMed controller tools
System manual MeDis 5MP1HM 75Figure 33: Windows 2000 Screen Saver TabFigure 34: Windows XP Screen Saver TabBarcoMed controller tools
System manual MeDis 5MP1HM 763. Click on the Settings button. The BARCO DPMS Screen SaverControl Panel will open (see figure 35).Figure 35: BARCO
System manual MeDis 5MP1HM 77Figure 36To create a new scheme by modifying one of the three defaultschemes please do the following.1. Open the Display
System manual MeDis 5MP1HM 7810. Save your preferences by pressing the Save As button andentering a new profile name in the Save Profile Setting d
System manual MeDis 5MP1HM 793. Select the radio button next to Remove on the ProgramMaintenance screen of the Wizard and click Next (figure 38).F
System manual MeDis 5MP1HM 8(This page intentionally left blank.)Introduction
System manual MeDis 5MP1HM 80Figure 406. Click Quit to exit the BarcoMed Product Installation Wizard.BarcoMed controller tools
System manual MeDis 5MP1HM 81MEDICAL SOFTWAREINSTALLATION ANDUSAGE
System manual MeDis 5MP1HM 82(This page intentionally left blank.)MediCal software installation and usage
System manual MeDis 5MP1HM 83MEDICAL SOFTWARE INSTALLATIONInstall MediCal on the PC. Follow the instructions from the MediCalUser Guide. Important: Do
System manual MeDis 5MP1HM 84(This page intentionally left blank.)MediCal software installation and usage
System manual MeDis 5MP1HM 85DISPLAY USERMANUAL
System manual MeDis 5MP1HM 86(This page intentionally left blank.)Display User Manual
System manual MeDis 5MP1HM 87© 2003 BARCO nv. All rights reserved.Trademark informationTrueGrey, MediCal, BarcoMed, and MeDis are registered trademark
System manual MeDis 5MP1HM 88Regulations This apparatus conforms to: IEC601-1,UL2601-1, cUL2601-1, EN 60601-1 This apparatus is classifie
System manual MeDis 5MP1HM 89INTRODUCTIONOverviewResolution and bandwidthsThe BARCO MGD 521 M and MGD 521 MKII displays are ultra-highresolution, gray
System manual MeDis 5MP1HM 9INTRODUCTIONBARCO MeDis® 5MP1H is a total display solution for medicalapplications. This can be a single-head or multi-hea
System manual MeDis 5MP1HM 90adjust and check BARCOs medical displays.A lot of adjustments can also be done on the monitor itself, by meansof an exte
System manual MeDis 5MP1HM 91About the manualsThis guide is meant for people who want to install and use the MGD521 displays, as well as people who ne
System manual MeDis 5MP1HM 92INSTALLATIONPrecautions Keep your original packaging. It is designed for this display and isthe ideal protection during
System manual MeDis 5MP1HM 93Controls and connectors1REMOTE OUTREMOTE INVIDEOHS / CSVS245678910113{1} Optical sensor plug{2} Ambient Light Compensatio
System manual MeDis 5MP1HM 94Signal connectiona) Connection of video signalsThe connected equipment must comply to all relevant safetydemands.Importan
System manual MeDis 5MP1HM 95b) Connection of data signalsThe display can be controlled remotely by a computer through theserial data bus. A typical e
System manual MeDis 5MP1HM 96d) Connection of power1. Power cord: Utilize a UL-listed detachable power cord, 3-wire, typeSJ or equivalent, 18 AWG min.
System manual MeDis 5MP1HM 97Important considerations The best environment for diagnostic imaging is one withcontrolled and dimmed ambient light. The
System manual MeDis 5MP1HM 98OPERATION: USER CONTROLSImportant:The best way to adjust and control the display is by using MediCal,which displays the i
System manual MeDis 5MP1HM 99Contrast and brightness controlContrast and brightness can be controlled through the OSD menus.To activate the OSDTurn th